Clinical Evaluation of A-LIKNet: Deep Learning-accelerated Single-breath-hold CINE Magnetic Resonance Imaging for Cardiac Function Assessment.

Journal: Investigative radiology
Published Date:

Abstract

OBJECTIVES: To evaluate the diagnostic accuracy and quantitative agreement of A-LIKNet (attention-incorporated network for sharing low-rank, image, and k-space information) deep learning (DL)-accelerated 2D cardiac CINE MRI acquired in a single breath-hold, compared with standard multi-breath-hold CINE sequences, for assessing biventricular volumes and function. MATERIALS AND METHODS: In this single-center study, A-LIKNet DL-reconstructed CINE images were acquired at 3 acceleration factors (8×, 16×, and 24×) in 42 subjects using a single breath-hold protocol. Quantitative parameters, including left and right ventricular end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF), and left ventricle myocardial mass (MM), were derived from standardized segmentation workflows. These were compared against standard multi-breath-hold CINE sequences and tested for agreement using Bland-Altman analysis and linear regression. Image quality was assessed using the coefficient of variation (CoV). RESULTS: Excellent agreement was observed between A-LIKNet DL-accelerated and standard CINE imaging for EDV, ESV, SV, and EF, with 95% limits of agreement (LoA) falling within predefined equivalence margins for nearly all parameters. No significant proportional bias was detected. Myocardial mass showed wider variability and exceeded equivalence thresholds, likely due to less distinct epicardial borders in higher acceleration. CoV values increased with higher acceleration, reflecting mild degradation in image quality, although segmentation performance remained robust across all levels. CONCLUSION: A-LIKNet deep learning accelerated CINE sequences enable rapid and reliable assessment of cardiac function with excellent agreement to standard imaging. From a clinical perspective, image quality remains acceptable up to an acceleration factor of 16×, supporting routine application, while 24× acceleration may be reserved for selected use cases requiring maximal speed.

Authors

  • Jens Kübler
    Department of Radiology, Diagnostic and Interventional Radiology, University of Tübingen, 72076 Germany.
  • Stanislau Chekan
  • Siying Xu
    Medical Image and Data Analysis (MIDAS.lab), Department of Diagnostic and Interventional Radiology, University of Tuebingen, Tuebingen, Germany. Electronic address: [email protected].
  • Aya Ghoul
  • Andreas Lingg
    From the Department of Diagnostic and Interventional Radiology, Eberhard Karls University Tuebingen, Tuebingen.
  • Florian Hagen
    Department of Diagnostic and Interventional Radiology, Eberhard-Karls-University, Hoppe-Seyler-Str.3, 72076 Tübingen, Germany.
  • Jan-Michael Brendel
  • Konstantin Nikolaou
    Department of Diagnostic and Interventional Radiology, Eberhard-Karls-University, Hoppe-Seyler-Str.3, 72076 Tübingen, Germany.
  • Thomas Küstner
    Department of Radiology, Diagnostic and Interventional Radiology, Eberhard Karls University Tübingen, Germany.
  • Patrick Krumm
    Department of Radiology, Diagnostic and Interventional Radiology, University of Tübingen, 72076 Germany. Electronic address: [email protected].

Keywords

No keywords available for this article.