Dual-Domain Sparse-View 4D CBCT Reconstruction with Spatial-Temporal Consistency Learning
Journal:
arXiv
Published Date:
Oct 13, 2026
Abstract
Four-dimensional cone-beam computed tomography (4D CBCT) provides phase-resolved volumetric images for image-guided radiotherapy. Yet good reconstruction quality usually needs many projections. This increases scan time and radiation dose. We propose STCNet, a sparse-view 4D CBCT reconstruction framework that uses spatial-temporal consistency learning. STCNet reconstructs high-quality images from only 16 projections in each respiratory phase. STCNet integrates physics-driven positional encoding to enforce geometric consistency in the projection domain and introduces temporal encoding in latent space to capture respiratory motion dynamics across phases to capture dynamic respiratory motion across phases. A dual-domain collaborative strategy jointly optimizes both projection and image domains, enhancing structural fidelity and temporal coherence. We tested STCNet on the TCIA-LUNG dataset and a CIRS thorax phantom dataset. STCNet performed better than conventional methods and recent deep learning baselines in image quality, artifact reduction, and structural preservation. On TCIA-LUNG, STCNet reached PSNR values of 33.311, 32.695, and 32.658 dB at phases 2, 5, and 8. The corresponding SSIM values were 0.845, 0.823, and 0.825. On CIRS, STCNet achieved PSNR values up to 36.497 dB and SSIM values up to 0.877. These results support STCNet as a promising framework for low-dose, phase-resolved 4D CBCT reconstruction in image-guided radiotherapy.