[Determination of Plasticizer DEHP Released from Medical Injection Equipment].

Journal: Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
PMID:

Abstract

To established an effective GC-MS /MS method for the contents determination of the residual DEHP in injection equipment, and investigate the effect of the pretreatment on the measurement. To simulate the clinical conditions of use, under the condition of 37 degrees C balance extraction, extract liquor by chloroform extraction, then the extract followed by analysis of GC-MS /MS. The method was simple, rapid, sensitive and accurate. The limits of quantitation (LOQ, S/N = 5) of cyclohexanone was 0.075 μg/mL, The spiked average recoveries ranged from 92% to 98%. The relative standard deviations (RSDs) of the method ranged from 1.01% to 1.61%, The method was simple, fast, sensitive and accurate, and may serve as a mass control method for residual DEHP in injection equipment.

Authors

  • Jian Zheng
    Biospheric Assessment for Waste Disposal Team, National Institute of Radiological Sciences, National Institutes for Quantum and Radiological Science and Technology, 4-9-1 Anagawa, Inage, Chiba 263-8555, Japan; Fukushima Project Headquarters, National Institute of Radiological Sciences, National Institutes for Quantum and Radiological Science and Technology, 4-9-1 Anagawa, Inage, Chiba 263-8555, Japan. Electronic address: zheng.jian@qst.go.jp.
  • Minzhu Wang
  • Wang Anyan
  • Pinghua Xu
  • Yin Han
  • Yan Wen
  • Li Zhang
    Department of Animal Nutrition and Feed Science, College of Animal Science and Technology, Huazhong Agricultural University, Wuhan 430070, China.