[Realization and Clinical Study of Robotic Technology in Cardiovascular Interventional Surgery].

Journal: Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Published Date:

Abstract

The high incidence of cardiovascular diseases is a serious threat to human health, and endovascular surgery has become the standard treatment for most interventional cardiovascular diseases. The robotassisted endovascular surgery system further enhances surgeons' ability to perform minimally invasive endovascular procedures in interventional cardiology. This study presents a new robotic technique for coronary intervention from the perspective of clinical application. Aiming at clinical application scenarios, this scheme proposed an intuitive guide wire catheter mechanism design, which accurately and perfectly simulates the doctor's hand movements, realizes the positive and negative direction translation of the guide wire catheter, accurate torque control of the guide wire rotation and locking. The results of animal test showed that the R-One has a high degree of dexterity, accuracy and stability,and meets the clinical needs.

Authors

  • Shu-Xiang Li
    Shanghai Microport Medbot (Group) Co. Ltd., Shanghai, 201203.
  • Wen-Zhi Liu
    Cathbot (Shanghai) Robot Co. Ltd., Shanghai, 201203.
  • Hao-Yang Mao
    Shanghai Microport Medbot (Group) Co. Ltd., Shanghai, 201203.
  • Zhen-Qi Zhou
    Cathbot (Shanghai) Robot Co. Ltd., Shanghai, 201203.
  • Zhong-Lu Wang
    Cathbot (Shanghai) Robot Co. Ltd., Shanghai, 201203.
  • Jian-Bo Qiu
    Cathbot (Shanghai) Robot Co. Ltd., Shanghai, 201203.
  • Xu Jin
    School of Science, Nanjing University of Posts and Telecommunications, Nanjing 210046, China.