A semi-autonomous robot control based on bone layer transition detection for a safe pedicle tapping.
Journal:
International journal of computer assisted radiology and surgery
PMID:
36877289
Abstract
PURPOSE: Automatic robotic platforms for robot-aided spinal surgery are mostly employed for drilling the pedicle screw path and do not adapt the tool rotational speed depending on the variation of the bone density. This feature is highly desirable in control strategies for robot-aided pedicle tapping, which may result in a poor quality thread if the surgical tool speed is not adequately tuned depending on the bone density to be threaded. Therefore, the objective of this paper is to propose a novel semi-autonomous control for robot-aided pedicle tapping that is able to (i) identify the bone layer transition, (ii) adapt the tool velocity depending on the detected bone layer density and (iii) stop the tool tip before propulsion of the bone boundaries.