[Evaluation on Surgical Robotic Pose Repeatability Based on Laser Tracking Measurement].

Journal: Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation
Published Date:

Abstract

To provide accurate information for registration and safety evaluation of surgical robot, the pose repeatability measurement method was proposed. According to the terminal instrument of the master-slave surgical robot (such as high-frequency electric knife, ultrasonic knife), a suitable target ball fixture was designed. The node data at 10%, 50% and 100% rated speed were measured respectively. Through data analysis, the pose repeatability property of the tested samples at different speeds was obtained. It has high applicability and repeatability, and can meet the requirements of data traceability and registration testing.

Authors

  • Jian Jin
    Department of Agricultural and Biological Engineering, Purdue University, 225 S. University St., West Lafayette, IN 47907, USA.
  • Zhengjie Sun
    Zhejiang Institute of Medical Device Supervision and Testing, Hangzhou, 310018.
  • Kun Du
    Zhejiang Institute of Medical Device Supervision and Testing, Hangzhou, 310018.
  • Wei Chen
    Department of Urology, Zigong Fourth People's Hospital, Sichuan, China.