PLS-LS-SVM based modeling of ATR-IR as a robust method in detection and qualification of alprazolam.
Journal:
Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy
Published Date:
Feb 15, 2017
Abstract
According to the United States pharmacopeia (USP), Gold standard technique for Alprazolam determination in dosage forms is HPLC, an expensive and time-consuming method that is not easy to approach. In this study chemometrics assisted ATR-IR was introduced as an alternative method that produce similar results in fewer time and energy consumed manner. Fifty-eight samples containing different concentrations of commercial alprazolam were evaluated by HPLC and ATR-IR method. A preprocessing approach was applied to convert raw data obtained from ATR-IR spectra to normal matrix. Finally, a relationship between alprazolam concentrations achieved by HPLC and ATR-IR data was established using PLS-LS-SVM (partial least squares least squares support vector machines). Consequently, validity of the method was verified to yield a model with low error values (root mean square error of cross validation equal to 0.98). The model was able to predict about 99% of the samples according to R of prediction set. Response permutation test was also applied to affirm that the model was not assessed by chance correlations. At conclusion, ATR-IR can be a reliable method in manufacturing process in detection and qualification of alprazolam content.