The Delta Robot-A long travel nano-positioning stage for scanning x-ray microscopy.
Journal:
The Review of scientific instruments
Published Date:
Apr 1, 2022
Abstract
A new stage design concept, the Delta Robot, is presented, which is a parallel kinematic design for scanning x-ray microscopy applications. The stage employs three orthogonal voice coils, which actuate parallelogram flexures. The design has a 3 mm travel range and achieves rms position jitter, integrated from 1 Hz to 1 kHz, of 2.8 and 1.3 nm perpendicular to the beam and 5.6 nm along the beam direction with loads up to 350 g. The Delta Robot design process used a mechatronics approach of iterative modeling and simulation to develop the system and validate performance. The design considerations, design process, stability, and operational performance on the hard x-ray nanoprobe at Diamond Light Source are presented.