Dosimetric study on learning-based cone-beam CT correction in adaptive radiation therapy.
Journal:
Medical dosimetry : official journal of the American Association of Medical Dosimetrists
Published Date:
Apr 1, 2019
Abstract
INTRODUCTION: Cone-beam CT (CBCT) image quality is important for its quantitative analysis in adaptive radiation therapy. However, due to severe artifacts, the CBCTs are primarily used for verifying patient setup only so far. We have developed a learning-based image quality improvement method which could provide CBCTs with image quality comparable to planning CTs (pCTs). The accuracy of dose calculations based on these CBCTs is unknown. In this study, we aim to investigate the dosimetric accuracy of our corrected CBCT (CCBCT) in brain stereotactic radiosurgery (SRS) and pelvic radiotherapy.